| Model Number |
3389S-40 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that high lead impedance was detected during the deep brain stimulation procedure. the surgery was completed on the same day after replacing the leads.No patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis # (b)(4) :analysis information -- 2026-02-05 13:15:00 cst pli# 20 product id# 3389s-40 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The returned device passed all testing in the laboratory and no anomalies were identified.Continuation of d10: product id: neu_stylet_acc lot# serial# unknown.Product type ; accessory product id neu_stylet_acc lot# serial# unknown.Product type ; accessory product id 3389s-40 lot# va323cq serial# implanted: explanted: product type ; lead.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|