|
The event is rated as confirmed, since the visible damages indicate that there were multiple issue during the screw insertion.The device inspection revealed the following: one cancellous bone screw was returned without allegation, but during the visual inspection it was detected that the screw is strongly damaged.The first three thread flanks are totally flattened; this indicates that there was an very strong contact with another implant at the end of the screw insertion.In addition, there is a deep notch running across the thread on the shaft.This notch was most likely caused by the drill bit during pre-drilling for another screw.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.The event was caused by selecting a screw trajectory, with a path that led to a collision with other implants.If this event had an effect on the locking screws can in retrospective not be defined.If more information is provided, the case will be reassessed.
|