Catalog Number 1758SI16
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.All available udi information is being provided.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that in or department, inserted one of these foley catheters in a patient and could not get the balloon to inflate.They opened a 2nd tray and had the same issue with that tray.They did not keep any of the used catheters.They only have the outer packages, but they do not have the lot numbers on the outer packaging.Per customer follow up information received via email on 17nov2025, it was reported that they have the outer wrappers for the 2 foley trays, but unfortunately the staff did not retain the defective catheters.Lot number was not available.Two urinary catheter insertions were required.Removal and reinsertion of urinary catheter were performed.They were performed troubleshooting.Attempted to insert a foley and the balloon would not inflate.Foley removed.Opened another foley kit and while testing the balloon a leak was seen, and this balloon would also not inflate.Opened another foley kit, balloon tested and inflated appropriately.Foley inserted and then unable to draw urine from the port with syringe.
 
Event or Problem Description
It was reported that in or department, inserted one of these foley catheters in a patient and could not get the balloon to inflate.They opened a 2nd tray and had the same issue with that tray.They did not keep any of the used catheters.They only have the outer packages, but they do not have the lot numbers on the outer packaging.Per customer follow up information received via email on 17nov2025, it was reported that they have the outer wrappers for the 2 foley trays, but unfortunately the staff did not retain the defective catheters.Lot number was not available.Two urinary catheter insertions were required.Removal and reinsertion of urinary catheter were performed.They were performed troubleshooting.Attempted to insert a foley and the balloon would not inflate.Foley removed.Opened another foley kit and while testing the balloon a leak was seen, and this balloon would also not inflate.Opened another foley kit, balloon tested and inflated appropriately.Foley inserted and then unable to draw urine from the port with syringe.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A dhr could not be performed since no lot number was provided.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® I.C. ALL-SILICONE COATED 5 CC FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23639922
Report Number1018233-2025-10523
Device Sequence Number9841102
Product Code EZL
UDI-Device Identifier00801741025228
UDI-Public(01)00801741025228
Combination Product (Y/N)Y
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1758SI16
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2025
Supplement Date Received by Manufacturer04/20/2026
Initial Report FDA Received Date11/24/2025
Supplement Report FDA Received Date04/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age78 YR
Patient SexFemale
Patient Weight80 KG
Patient RaceWhite
-
-