• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO BIPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO BIPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX700S11
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted as part of a batch submission of complaints deemed reportable following record correction and remediation efforts.
 
Event or Problem Description
A device was returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service centre, the device was visually inspected and found evidence of foam degradation.During the evaluation, secondary findings was observed.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service centre.There was no error code found.The third-party service center concludes that there was visible foam degradation and unit got corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO BIPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key23643510
Report Number2518422-2025-056528
Device Sequence Number10133171
Product Code BZD
UDI-Device Identifier00606959424043
UDI-Public(01)00606959424043
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 11/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberDSX700S11
Device Catalogue NumberDSX700S11
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2023
Initial Report FDA Received Date11/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/04/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-