| Catalog Number |
IAB-06830-U |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported " after the catheter into the patient, the pump alarmed helium loss 3.After the catheter was removed, it was found that the front end of the balloon was bent".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned for investigation was a 30cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) without the original packaging.The sample was returned in a cardboard box and was in a sealed ziploc bag.Up-on return, the teflon sheath was noted on the iabc outer lumen.The one-way valve was tethered to the short driveline tubing.The bladder was fully unwrapped.The iabc central lumen was noted dam-aged, consistent with being broken at approximately 5.1cm from the iabc distal tip.Dried blood was noted on the exterior surfaces of the returned sample.No obvious blood was noted within the helium pathway.No other visual damage or abnormalities were noted to the returned sample.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.An attempt to aspirate and flush the iabc central lumen using a 60cc lab-inventory syringe was unable to be successfully completed.Upon aspiration, water was unable to be consistently pulled into the syringe.The attempt to flush resulted in the blad-der inflation, which is consistent with the previously confirmed broken central lumen.The iabc was leak tested in accordance with testing methods from manufacturing procedure.A leak was immediately detected from the iabc distal tip and iabc luer end.The leak from the iabc distal tip and iabc luer end are consistent with the previously confirmed broken central lumen.The iabc was leak tested again with the iabc distal tip and iabc luer end blocked off; no other leaks were detect-ed.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.The guidewire could not advance at approximately 5.1cm from the iabc distal tip, which is the location of the pre-viously confirmed damaged central lumen.No blood or debris was noted.The guidewire was front loaded through the iabc luer.The guidewire could not advance at approximately 72.1cm from the iabc luer, which is the location of the previously confirmed damaged central lumen.No blood or de-bris was noted.A device history record (dhr) r eview was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint that the "pump alarmed helium loss 3" is confirmed.During the investigation, the intra-aortic balloon catheter (iabc) central lumen was noted broken near the iabc distal tip.A broken central lumen can cause a helium loss alarm.Based on a review of the device history rec-ord (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the broken central lumen.The root cause of the complaint is undetermined.No further action required at this time.This will be monitored for any developing trends.
|
| |
|
Event or Problem Description
|
|
It was reported " after the catheter into the patient, the pump alarmed helium loss 3.After the catheter was removed, it was found that the front end of the balloon was bent".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Search Alerts/Recalls
|