| Catalog Number |
8065751761 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Excessive Heating (4030)
|
| Health Effect - Clinical Codes |
Wound Dehiscence (1154); Blurred Vision (2137); Eye Burn (2523)
|
| Date of Event |
11/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
The nurse reported that the female patient during cataract surgery on the right eye ophthalmic handpiece phaco tip got very hot and the patient experienced corneal burn which results in would leakage.The patient had suturing as surgical intervention.The surgery was completed on the same day.The current condition of the patient was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections h.6 and h.11.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.There was no product returned for this investigation.Based on the information obtained, the root cause of the reported event is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.No similar complaints were reported for the product serial (under investigation), with the same event code.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|