• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS INC. SURFACER® INSIDE-OUT®; REVERSE CENTRAL VENOUS RECANALIZATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS INC. SURFACER® INSIDE-OUT®; REVERSE CENTRAL VENOUS RECANALIZATION SYSTEM Back to Search Results
Catalog Number 600200/C
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Hemothorax (1896)
Date of Event 08/07/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect medical device has not been returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.No lot number was provided; thus, the device history record could not be reviewed and a search of the complaint database could not be performed.
 
Event or Problem Description
In a literature article (bohle et al., 2023), customer reports "a 71-year-old female undergoing surfacer inside-out central venous catheter placement for thoracic central venous occlusion experienced a near-fatal arterial injury.During advancement of the device, the right internal mammary artery was inadvertently lacerated, resulting in massive bleeding and cardiovascular collapse.The patient required transfusion, chest tube placement, and coil embolization of the injured artery.Imaging review post-procedure revealed the trajectory of the device intersected the artery, indicating potential for injury due to anatomical variation.The event resulted in hemothorax and prolonged hospitalization, but the patient ultimately recovered.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURFACER® INSIDE-OUT®
Common Device Name
REVERSE CENTRAL VENOUS RECANALIZATION SYSTEM
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan UT 84095
Manufacturer Contact
adrian cole
1600 merit parkway
south jordan, UT 84095
MDR Report Key23645271
Report Number1721504-2025-00623
Device Sequence Number9829840
Product Code QJH
UDI-Device Identifier00884450762158
UDI-Public00884450762158
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Literature
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue Number600200/C
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2025
Initial Report FDA Received Date11/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-