| Catalog Number |
02.221.092 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional manufacturer narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that the insertion handle for the ppfx plating system would not thread into plates.As well as messed up threads on plate.It was not known if fragments were generated.It was unknown if surgery was delayed due to the reported event and if procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d4 (lot, expiry date), d9 corrected: d1, d4 (primary udi number) the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: h4.Corrected: g1 (manufacturing site name).H3, h4, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that one threaded hole of the va-lcp ppfx proximal femur hook plate 3.Were heavily stripped.No other defects were found.The observed condition of the device is consistent with a component failure that was caused by exposure to unintended forces like overtightening the device while assemblance; or by assembling the device in a misaligned position.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional evaluation was not performed due to the mating device was not returned for evaluation.The overall complaint was confirmed as the observed condition of the va-lcp ppfx proximal femur hook plate 3.Would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part number: 38204p4.Lot number: 02.221.092.Manufacturing site: mezzovico.Release to warehouse date: 25 oct 2024.A manufacturing record evaluation was performed for the not sterile finished lot number, and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|