| Catalog Number |
EL31915 |
| Medical Device Problem Code |
Partial Blockage (1065)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the syringe became blocked when the patient was injected.There was reported patient involvement, but no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product was returned; therefore, no visual and functional tests were performed, the reported complaint could not be confirmed, and the root cause could not be determined.If the product is returned, the manufacturer will reopen this complaint for further investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: possible lot numbers: 25a014, 6002494, 4279250, 6064770.
|
| |
|