|
A patient underwent a 4-level anterior cervical discectomy and fusion procedure on (b)(6) 2024.Around one year postoperatively, a radiographic image revealed that the locking mechanism had broken off, causing the screw to backout and fracture.The patient is asymptomatic, and there are no plans for revision surgery.
|
|
The implant has not returned for evaluation as it remains in the patient.A radiographic image confirms the event of a backout and fracture of the most superior screw.A review of device history records could not be performed as the lot number was not provided.Based on the information provided, the root cause could not be determined.Alphatec spine is submitting this report to comply with the fda regulations 21 cfr 803.Alphatec spine has made reasonable efforts to provide as much relevant information as available.Any field that is left blank was not known at the time of this submission.If additional information is provided, a supplemental report will be submitted.Warnings/cautions/precautions: without solid bone fusion, this device cannot be expected to support the cervical spine indefinitely and may fail due to bone-metal interface, metal, or bone failure.Based on the fatigue testing results, the physician/surgeon should consider the levels of implantation, patient weight, patient activity level, compliance of the patient, and other patient conditions which may have an impact on the performance and results of this system.No spinal implant can withstand body loads for an indefinite period of time without the support of bone.In the event that successful fusion is not achieved; bending, breakage, loosening, migration and/or disassembly of the device will occur.Potential risks identified with the use of this device, which may require additional surgery, include device component fracture, loss of fixation, non-union, fracture of the vertebrae, and necrosis of bone, neurological injury and vascular or visceral injury.Possible adverse effects: the following complications and adverse reactions have been shown to occur with the use of similar spinal instrumentation.These effects and any other known by the surgeon should be discussed with the patient preoperatively.Initial or delayed loosening, disassembly, bending, dislocation and/or breakage of device components.Postoperative management: the patient should have a complete understanding of and compliance with the purpose and limitations of the implant devices.The surgeon should instruct the patient on how to compensate for any loss in range of spinal motion due to bone fusion.Additional or revision surgery may be necessary to correct an adverse effect.The surgeon should instruct the patient regarding amount and time frame after surgery of any weight bearing activity.The increased risk of bending, dislocation, and /or breakage of the implant devices, as well as an undesired surgical result are consequences of any type of early or excessive weight bearing, vibration motion, fall, jolts or other movements preventing proper healing and/or fusion development.
|