| Model Number |
Ø9 X 30MM PEEK IF SCRW NON-VENTED |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament reconstruction the screw did not hold in the drill channel.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a different device (ar-4030p-10).It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.-complaint allegation is confirmed.-visual inspection of the fastthread¿ peek interference screw, 9x30mm identified that the threads were damaged.-functional testing was not performed due to the damage to the device.The method of bone preparation and the quality of the bone encountered were not specified.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misalignment insertion, and/or prying/leveraging of the device during insertion.
|
| |
|