ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported multiple tasp shims were returned missing ball bearings on a worn instrument claim form.It was further clarified that all those instruments were from long time ago, and they are not from the same procedure and had no specific reasons for the issues.Attempts were made and all available information was provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h4; h6; h9.Visual examination of the returned tasp identified signs of use (scuffs, dings) and confirmed the complaint, one of the bearings and springs was disassembled / missing and not returned.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23649170
Report Number0001822565-2025-04302
Device Sequence Number10123850
Product Code MBH
UDI-Device Identifier00889024245952
UDI-Public(01)00889024245952(10)64451594
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42527900701
Device Lot Number64451594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2025
Supplement Date Received by Manufacturer04/22/2026
Initial Report FDA Received Date11/25/2025
Supplement Report FDA Received Date04/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-