This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: investigation summary the product was not returned to depuy synthes, however a photo was provided for evaluation.¿ visual inspection of the returned photo found that the screw was still attached to the inserter and was broken near mid-body.Additionally, the device had foreign matter, presumably biological matter.No other anomalies were noted.¿ the overall complaint was confirmed as the observed condition of the milagro advance screw 7x23mm would have contributed to the complained device issue.¿ ¿ based on the investigation findings, the potential cause is traced to the procedural variables, such handling of the device or product interaction during procedure.Resistance may have been felt when the device was being inserted and the excessive force applied caused the screw to break and deform.As per the instructions for use (ifu); 1) if resistance is felt during the insertion of a milagro br interference screw over a guidewire, stop and confirm that the guidewire is not entrapped.If entrapped, back out the screw and withdraw the guidewire.Properly handling and¿attention to the approved use of the device diminishes the risk of failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.¿.
|