XERIDIEM MEDICAL DEVICES ENDOVIVE GASTROSTOMY TUBE, RIGHT ANGLE, WITH ENFIT CONNECTOR; ENDOVIVE G-TUBES WITH ENFIT CONNECTOR
|
Back to Search Results |
|
| Model Number |
70-0053-222 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Burn(s) (1757); Inflammation (1932); Discomfort (2330); Gastrointestinal Hemorrhage (4476); Swelling/ Edema (4577)
|
| Date of Event |
10/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint number: (b)(4).Root cause summary: no manufacturing defect was identified.The failure is likely due to external mechanical, chemical, or environmental factors beyond the manufacturer's control.Sample evaluation: devices were not returned for investigation.Photos depicting complaint symptoms were provided by the customer.These photos confirm the defect; some of the devices that were in the photos are not manufactured by xeridiem.
|
| |
|
Event or Problem Description
|
|
Ecn event description: situación que nos reportan en sura de las sondas de recambio enfit, han tenido más de 5 pctes, los cuales han presentado sangrado periostomal desde el cambio de las sondas, filtraciones, daño del estoma y piel perilesional, esto no había ocurrido antes y se ha venido evidenciando desde las colocaciones de septiembre, agradezco con latam si puedes validar si han presentado una situación similar.Comparto evidencias fotográficas de los casos situation reported to us at sura regarding enfit replacement tubes: they have had more than 5 patients who have presented peristomal bleeding since the tube change, leaks, damage to the stoma and perilesional skin.This had not occurred before and has been evident since the september placements.I would appreciate it if latam could validate whether they have presented a similar situation.I share photographic evidence of the cases what troubleshooting steps took place? what troubleshooting steps, if any, resolved the issue?: valoración por especialista - specialist assessment what is the next course of action?: valoración por especialista - specialist assessment patient present at time of event?: yes patient complications: discomfort / pain,edema / inflammation / swelling,fluid discharge,burn(s) in the physician's opinion, did the device or procedure cause or contribute to the patient complication?: unknown medical or surgical interventions: unknown patient admitted to hospital beyond the standard of care: unknown patient outcome: the issue is ongoing device acquired from a distributor?: no.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint number: (b)(4).There were two lots identified in the complaint description: lot: 11521-11 with expiry date - 04-04-2027.Lot: 11490-09 with expiry date - 02-23-2027.Root cause summary: no manufacturing defect was identified.The failure is likely due to external mechanical, chemical, or environmental factors beyond the manufacturer's control.Sample evaluation: devices were not returned for investigation.Photos depicting complaint symptoms were provided by the customer.These photos confirm the defect; some of the devices that were in the photos are not manufactured by xeridiem.Corrections/additions from initial report: added the lot numbers for the devices, expiration date and manufacturing date.This information was provide in a follow-up email from the distributor.It was not included in the initial complaint received by the manufacture.
|
| |
|
Event or Problem Description
|
|
Ecn event description: situaci6n que nos reportan en sura de las sondas de recambio enfit, han tenido mas de 5 pctes, las cuales han presentado sangrado periostomal desde el cambio de las sondas, filtraciones, dai'io del estoma y piel perilesional, esto no habia ocurrido antes y se ha venido evidenciando desde las colocaciones de septiembre, agradezco con latam si puedes validar si han presentado una situaci6n similar.Comparto evidencias fotograficas de las cases situation reported to us at sura regarding enfit replacement tubes: they have had more than 5 patients who have presented peristomal bleeding since the tube change, leaks, damage to the stoma and perilesional skin.This had not occurred before and has been evident since the september placements.I would appreciate it if latam could validate whether they have presented a similar situation.I share photographic evidence of the cases what troubleshooting steps took place? what troubleshooting steps, if any, resolved the issue? valoraci6n par especialista - specialist assessment what is the next course of action? valoraci6n par especialista - specialist assessment patient present at time of event? yes patient complications: discomfort/ pain.Edema/ inflammation / swelling.Fluid discharge, burn(s) in the physician's opinion, did the device or procedure cause or contribute to the patient complication?: unknown medical or surgical interventions: unknown patient admitted to hospital beyond the standard of care: unknown patient outcome: the issue is ongoing device acquired from a distributor? no.
|
| |
|