| Model Number |
60005 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
10/29/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A site reported to rxsight that a patient's light adjustable lens (lal) was explanted due to the patient's report of peripheral vision loss after lock in.Per the attending physician, the patient's visual field was normal.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer reference#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information to the following sections: d4, d6a, h4, h6, and h11.The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.Manufacturer reference #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h6 (component code)
additional data: d9, h3, h6, (type of investigation, investigation findings), h11
the explanted lal was returned for evaluation.A visual inspection was performed and the lens was observed to have been cut into three main pieces during explantation.Additionally, the periphery of the lens optic was observed to be damaged which is likely due to the removal of the lens.However, this could not be confirmed without evidence of the lens' condition prior to explantation.The likely cause of the reported event was therefore unable to be identified.
|
| |
|