G2: foreign country - japan.H3,h6: the returned probe was analyzed.There was physical damage.The cable had been pulled exposing a small amount of the internal wires at the strain relief.No conductors were exposed.Otherwise, the probe displayed good geometry and divot error readings with normal tracking and verification.Codes b01, c07, and d02 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|