| Model Number |
DVBB1D4 |
| Medical Device Problem Codes |
Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient received high voltage therapy twenty-one times from the implantable cardioverter defibrillator (icd) that was deemed by the physician as inappropriate for supra ventricular tachycardia (svt).The physician noted that a magnet was applied to the device to interrupt the shocks.The icd was later explanted and replaced due to reaching normal elective replacement indicator (eri).No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.The battery indicator signifying that it is time for device replacement occurred after explant.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the device detected in the ventricular fibrillation (vf) zone.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|