MAUDE Adverse Event Report: BIOSENSE WEBSTER INC OCTARAY MAPPING CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC OCTARAY MAPPING CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY Back to Search Results
Catalog Number D160904
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with an octaray mapping catheter and the catheter stopped irrigating properly after it was used for a while and removed from the body.There was no visible damage to the catheter.When the catheter was replaced, the issue resolved.The faulty catheter was brand new and not reprocessed.There was no patient consequence.Additional information received indicated that only a pressurized heparinized saline bag was used, no automatic pump.There was no error code, and they only observed lack of flow.To resolve the irrigation issue, they increased the pressure in pressurized bag, they noticed a visual confirmation of irrigation flow from tubing.The same irrigation bag and tubing were used for the new octaray and irrigation was noticeably better.
 
Additional Manufacturer Narrative
The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Additional Manufacturer Narrative
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with an octaray mapping catheter and the catheter stopped irrigating properly after it was used for a while and removed from the body.Device evaluation details: the device was returned to johnson & johnson medtech (j&j) for evaluation.Visual inspection and irrigation test of the returned device were performed following j&j procedures.Visual analysis of the returned sample revealed no damage or anomalies on the device.An irrigation test was performed, and no issues were observed.The device was flushed, and no obstructed holes were observed during the test.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The irrigation issue reported by the customer could not be replicated during the product investigation.Other issues or circumstances may have occurred during the usage of the device that compromised its performance.The instructions for use (ifu) contain the following warnings and precautions: a compatible irrigation pump is recommended to ensure proper irrigation flow rate.Before use, verify that the irrigation ports are patent by infusing heparinized normal saline through the catheter and the irrigation tubing.Purge the catheter and the irrigation tubing with heparinized normal saline prior to insertion of the catheter into the patient.Inspect the irrigation saline for air bubbles prior to its use in the procedure.Air bubbles in the irrigation saline may cause emboli.Always maintain a constant heparinized normal saline infusion to prevent coagulation within the lumen of the catheter.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTARAY MAPPING CATHETER
Common Device Name
CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key23653058
Report Number2029046-2025-03916
Device Sequence Number20059335
Product Code MTD
UDI-Device Identifier10846835021233
UDI-Public(01)10846835021233(11)250814(17)280813(10)31740830L
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K193237
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD160904
Device Lot Number31740830L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2025
Supplement Date Received by Manufacturer12/06/2025
Initial Report FDA Received Date11/25/2025
Supplement Report FDA Received Date12/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FARAPULSE SYSTEM; NGEN¿ SYSTEM
Patient SexUnknown
-
-