• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRILLIANT SURGICAL, LLCD/BA DJO FOOT AND ANKLE STABILIZATION; 18MM X 15MM X 15MM STAPLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRILLIANT SURGICAL, LLCD/BA DJO FOOT AND ANKLE STABILIZATION; 18MM X 15MM X 15MM STAPLE Back to Search Results
Catalog Number 500-18-102
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/05/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - due to unknown reason.
 
Additional Manufacturer Narrative
See h6 and h11.The agent reported a revision surgery for unknown reason.Dhr review of ln 15173-04 was reviewed.Parts were released in march of 2020.The staple involved in this complaint event was implanted in the patient for an unknown length of time prior to removal, potentially up to five and a half years.There were no other sniper staple removal complaint events in the rolling year.Review of surgical technique would not be useful as limited information is available.The part was not returned and could not be inspected.Rf-001-0001 rev 10, sniper staple risk matrix, was reviewed.U4.0 relates to re-operation to remove or replace implant due to various reasons.The severity of 3, moderate, is appropriate as re-operation comes with inherent risk.The occurrence of 2, rare, is appropriate as this is the first reported case in the rolling year.The detection of 1, very high, is appropriate as the need for surgery would be very apparent.The risk matrix adequately captures this complaint event.There is very limited to no information available regarding the surgery to implant the original staple, the condition of the patient and implant during the period in which the staple was implanted, and the cause for removal surgery.Due to the limited information available, the root cause of the complaint event could not be narrowed down.The root cause will remain unknown at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STABILIZATION
Common Device Name
18MM X 15MM X 15MM STAPLE
Manufacturer (Section D)
TRILLIANT SURGICAL, LLCD/BA DJO FOOT AND ANKLE
727 north shepherd drive
suite 100
houston TX 77007
Manufacturer (Section G)
TRILLIANT SURGICAL, LLCD/BA DJO FOOT AND ANKLE
727 north shepherd drive
houston TX 77007
Manufacturer Contact
stacey bonnell
727 north shepherd drive
houston, TX 77007
8777883864
MDR Report Key23653537
Report Number3007420745-2025-00013
Device Sequence Number18672845
Product Code JDR
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
D309794
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number500-18-102
Device Lot Number5173-04(1)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2025
Supplement Date Received by Manufacturer04/24/2026
Initial Report FDA Received Date11/26/2025
Supplement Report FDA Received Date05/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-