| Catalog Number |
8065750597 |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported that an ophthalmic laser not producing any laser sound.The procedure details and patient impact were not reported.Additional information has been requested, but none received to date.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections b.3.D.9, h.3, h.6 and h.11.The company representative was able to replicate the reported event.The speaker cabling was replaced to address this issue.At the time of this investigation, servicing remains in the ¿opened state.¿ a non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.No similar complaints were reported for the product serial under investigation.The root cause of the reported event is attributed to a nonconforming speaker cabling.As to how or why the cabling became non-conforming, cannot be conclusively determined.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|