| Catalog Number |
0165L12 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that foley catheter was spontaneously removed 10 days after placement of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter name: (b)(6).The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed - cause unknown.The product had caused the reported failure.Visual inspection noted balloon burst along lumen without missing pieces.Introduced water into the inflation lumen and did not experience any obstructions to impede flow.Microscopic examination found there was present lubricant on the balloon area.However, the exact cause on how and when problem occurred could not be determined.The potential root cause for this failure could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/operator error).A review of the device labeling has been conducted for investigation purposes and was found adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action is required at this time.Corrections: d, f, h.Initial reporter name:(b)(6) medical center.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that foley catheter was spontaneously removed 10 days after placement of the catheter.
|
| |
|
Search Alerts/Recalls
|