| Catalog Number |
517770 / 90-9000KC |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.The device was not returned for evaluation.Without the benefit of analyzing the device, quality cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information the device was explanted and replaced due to tubing break.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes and inlet tubes of the pump.No functional abnormalities were noted with the pump.No functional abnormalities were noted with either cylinder 1 or cylinder 2.The surfaces of the detachment site of the inlet tubing of the reservoir appear to be rough and irregular, indicating stress was exerted.Based on examination of the returned product, it was concluded that while in-vivo both the exhaust tubes and inlet tube of the pump had overlapped and abraded against one another.While no functional abnormalities were noted during testing with the returned components, microscopic examination of the surfaces of the exhaust tube detachment end between the pump and reservoir revealed rough and irregular surfaces, indicating stress was exerted.Based on the overlapping of the pump tubes, in combination with device usage over time, this could contribute to sufficient stress to separate the exhaust tubing of the pump.A separation of this type would then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|