It was reported that during service and repair/pre-testing, it was observed that the attachment device heated up and was vibrating.The device also failed pretests for temperature verification and vibration assessment.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2025.All available information has been disclosed.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device the following steps failed: 2.3.3e temperature verification 2.3.3f vibration assessment during the service evaluation the following failures were identified: functional: heat functional: vibration conclusion: failure reported is confirmed.
|