| Model Number |
FIBERTAK DISPOSABLES STRAIGHT KIT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 11/6/2025, a sales representative reported via email that an ar-3638ds knotless fibertak disposable kit bound up and snapped at the hub near the drill during drilling.Additionally, an ar-3636 knotless 1.8 fibertak passing suture pulled out of the all-suture anchor, which appeared to be broken.The surgeon was able to remove all components from the patient.This was discovered during a labral repair procedure on (b)(6) 2025.Additional information was received on 11/13/2025: all fragments were successfully retrieved from the patient.The procedure was completed using an ar-3610pk-3 percutaneous insertion kit for the knotless fibertak anchor.A two-minute delay occurred while obtaining the replacement kit.No additional anesthesia was required.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided ¿ which may include the returned device (if available), photographs, videos, the event description, and any additional field input ¿ arthrex has determined the most likely cause.The most likely cause is attributed to user-related factors such as excessive user-applied mechanical forces, misaligned insertion, and improper bone preparation.
|
| |
|