U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FIBERTAK DISPOSABLES STRAIGHT KIT; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FIBERTAK DISPOSABLES STRAIGHT KIT; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number FIBERTAK DISPOSABLES STRAIGHT KIT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 11/6/2025, a sales representative reported via email that an ar-3638ds knotless fibertak disposable kit bound up and snapped at the hub near the drill during drilling.Additionally, an ar-3636 knotless 1.8 fibertak passing suture pulled out of the all-suture anchor, which appeared to be broken.The surgeon was able to remove all components from the patient.This was discovered during a labral repair procedure on (b)(6) 2025.Additional information was received on 11/13/2025: all fragments were successfully retrieved from the patient.The procedure was completed using an ar-3610pk-3 percutaneous insertion kit for the knotless fibertak anchor.A two-minute delay occurred while obtaining the replacement kit.No additional anesthesia was required.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Based on the information provided ¿ which may include the returned device (if available), photographs, videos, the event description, and any additional field input ¿ arthrex has determined the most likely cause.The most likely cause is attributed to user-related factors such as excessive user-applied mechanical forces, misaligned insertion, and improper bone preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERTAK DISPOSABLES STRAIGHT KIT
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23656704
Report Number1220246-2025-05209
Device Sequence Number17349418
Product Code LXH
UDI-Device Identifier00888867287440
UDI-Public(01)00888867287440(11)250721(17)300630(10)15469873
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFIBERTAK DISPOSABLES STRAIGHT KIT
Device Catalogue NumberAR-3638DS
Device Lot Number15469873
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2025
Supplement Date Received by Manufacturer03/05/2026
Initial Report FDA Received Date11/26/2025
Supplement Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-