| Catalog Number |
XMAX |
| Medical Device Problem Codes |
Leak/Splash (1354); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the motor device hose was damaged and was leaking a liquid.The device also failed pretests for hose assessment and oil leak assessment.It was reported in the service order that the device did not function.It was unknown when the event occurred.There were no reports of delays to a surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device according to n01.32a rev ad.The following steps failed: 4.3.3 hose assessment 4.3.14 oil leak assessment during the service evaluation the following failures were identified: functional: hose damage (excluding rupture) functional: leak - liquid conclusion: failure reported is not confirmed.
|
| |
|
Search Alerts/Recalls
|