| Catalog Number |
CON-HL-390 |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device displayed an over temperature alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for investigation.The device was functionally tested, but the investigation could not duplicate the reported issue.The investigation did identify that the device microswitch was broken.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Based on the results of the investigation, the reported complaint could not be confirmed.The device microswitch was replaced and the device passed all testing.
|
| |
|