Catalog Number CON-HL-390
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device displayed an over temperature alarm.
 
Additional Manufacturer Narrative
One device was received for investigation.The device was functionally tested, but the investigation could not duplicate the reported issue.The investigation did identify that the device microswitch was broken.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Based on the results of the investigation, the reported complaint could not be confirmed.The device microswitch was replaced and the device passed all testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Common Device Name
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ICU MEDICAL, INC.
201 west queen street
southington CT
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 60045
2247062300
MDR Report Key23657477
Report Number1526863-2025-00146
Device Sequence Number15983556
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public(01)10695085002796(11)260424
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K911383
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 11/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Initial Report FDA Received Date11/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/24/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-