This complaint is recorded by zimmer biomet under (b)(6).Review of the most recent repair record determined the motor speeds were unstable, the motor was corroded, the plug harness was damaged (no exposed metal wiring), the reciprocating arm and needle bearing were worn, and the device was out of calibration.The motor, plug harness, reciprocating arm, and needle bearing were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to component failure as the device exhibits wear and tear from repeated use.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor trends.G2: finland.
|