|
Additional information: d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-8692ds implant system, cpr¿ viper¿ with suturetape, white/blue, 1.3 mm, batch number 4113140677, was received for evaluation.Visual inspection found that the shaft of the meniscal viper and inserter were bent at the distal end close to the tip.The returned suture was found frayed.A functional test was not performed because the device was damaged.The most likely cause(s) of the reported failure are user error, misaligned insertion, or excessive forces applied by leveraging/prying the device.
|