| Model Number |
SGC0301 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2025 that the patient presented with grade 5 degenerative mitral regurgitation (mr) for a mitraclip procedure.During the steerable guide catheter(sgc) preparation, a leak was observed in hemostatic valve while flushing the guide.The sgc was replaced with another device to complete the procedure.The mr was reduced to grade <1 post procedure.There was no clinically significant delay or adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine a cause for the reported leak (loss of fluid column).There is no indication of a product issue with respect to manufacture, design, or labeling.
|
| |
|