HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D Back to Search Results
Model Number SDM-SYS-6000-3D
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
The user site reported that the c-arm of a selenia dimensions mammography systems, 3d system rotated on its own prior to a patient exam on (b)(6), 2025.No patient or user injuries were reported.A hologic field service engineer (fse) was dispatched to assess the system.The user site informed the fse that the c-arm of the system moved on its own when the rotary switch at the bottom of the c-arm was used.The fse observed that the rotary switch of the system in the back of the gantry was getting stuck, which caused the c-arm to move on its own.The fse ordered a new rotary switch for the system.The fse replaced the rotary switch and tested system and confirmed the switch was no longer getting stuck.The fse performed gain calibration and artifact evaluation and verified that the system is now operating according to manufacturer specifications.No further information is currently available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D
Common Device Name
MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
600 technology drive
newark DE 19702
Manufacturer (Section G)
HOLOGIC, INC
600 technology drive
newark DE 19702
Manufacturer Contact
dara smithers
600 technology dr
newark, DE 19702
MDR Report Key23658898
Report Number3005496266-2025-00101
Device Sequence Number18632338
Product Code OTE
UDI-Device Identifier15420045510760
UDI-Public(01)15420045510760(11)191204
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P080003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSDM-SYS-6000-3D
Device Catalogue NumberSDM-SYS-6000-3D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2025
Supplement Date Received by Manufacturer08/06/2026
Initial Report FDA Received Date11/26/2025
Supplement Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/04/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-