| Model Number |
MU-651RA |
| Medical Device Problem Code |
Unintended Application Program Shut Down (4032)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported on 11/04/2025 that this bedside monitor (bsm) was randomly rebooting.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer (bme) reported on 11/04/2025 that this bedside monitor (bsm) was randomly rebooting.No patient harm was reported.The bme said they move the bsm-1733's around and would not know which one was connected to the main bsm at the time the original issue occurred.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional device information: d10 concomitant medical device: the following device was used in conjunction with the bsm: bedside monitor (bsm-1733) model #: bsm-1733a serial #: (b)(6) device manufacturer data: ni unique identifier (udi) #: (b)(4).Returned to nihon kohden: na.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint: the biomedical engineer (bme) reported on (b)(6) 2025 that this bedside monitor (bsm) was randomly rebooting.No patient harm was reported.The bme said they move the bsm-1733's around and would not know which one was connected to the main bsm at the time the original issue occurred.Investigation summary: after the logs were reviewed by nihon kohden corporation (nkc); no logs indicating a reboot were found.However, error logs for the secure digital (sd) memory card were observed and recommended replacing the sd card.Trending will continue to be monitored for this device, incidents, issues, and causes.Additional device information: d10 concomitant medical device: the following device was used in conjunction with the bsm: bedside monitor (bsm-1733) model #: bsm-1733a serial #: ni device manufacturer data: ni unique identifier (udi) #: (b)(4) returned to nihon kohden: na.Additional information: b4 date of this report.G3 date received by manufacturer.G6 type of report.H2 if follow-up, what type? h6 event problem and evaluation codes.H11 additional manufacturer narrative.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported on (b)(6) 2025 that this bedside monitor (bsm) was randomly rebooting.No patient harm was reported.
|
| |
|
Search Alerts/Recalls
|