• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM; CATHETER INTRAVASCULAR, THERAPEUTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM; CATHETER INTRAVASCULAR, THERAPEUTIC Back to Search Results
Model Number 384539
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The sample is not available for return.The investigation is currently in progress.A follow-up report will be submitted upon investigation completion.The lot number provided "1163831" cannot be found in our system.
 
Event or Problem Description
A nurse reported "lcath midline leaking around purple hub within hours of placement and removed." a new picc was placed.There was no patient harm.
 
Additional Manufacturer Narrative
We conducted a thorough review of the manufacturing and inspection records for this lot and found no deviations or non-conformances.No samples were returned by the customer for evaluation.Additionally, the photographic evidence provided by the customer appeared to support the reported allegation.However, without access to the device affected, we are unable to complete a thorough investigation.As a result, we cannot determine a definitive root cause or establish an appropriate corrective action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM
Common Device Name
CATHETER INTRAVASCULAR, THERAPEUTIC
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek road
athens TX 75758
Manufacturer Contact
kelly mayer
1445 flat creek road
athens, TX 75758
9036759321
MDR Report Key23659191
Report Number0001625425-2025-01144
Device Sequence Number10055137
Product Code LJS
UDI-Device Identifier00886333209934
UDI-Public00886333209934
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K230669
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number384539
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2025
Supplement Date Received by Manufacturer04/08/2026
Initial Report FDA Received Date11/26/2025
Supplement Report FDA Received Date04/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-