| Model Number |
TH130 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not returned at the time of mdr submission.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the camera head failed during laparoscopy.No negative impact in state of health reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device in question was returned to manufacturer.This information is reflected in section d9.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation:
the device was sent to the manufacturer for inspection.During the manufacturer's technical inspection, the error description provided by the customer, "failed during laparoscopy", could be confirmed.The cause can be attributed to a random defect of the camera head cable.The shielding of the cable is broken and thus the hf radiation can influence the signal wires.Consequently, the image flickers or disappears during hf treatment.According to the ifu visual as well as functionality should always be checked before every use to avoid injuries and damages to the product.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|