Model Number CD2411-36C
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Further information was requested but not received.
 
Event or Problem Description
It was reported that the patient presented remotely via merlin.Net.Review of transmission revealed post-paced t-wave oversensing (pptwos) on the implantable cardioverter defibrillator (icd).Programming changes were discussed, no intervention was reported.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR ICD
Common Device Name
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key23659498
Report Number2017865-2025-1006122
Device Sequence Number10027480
Product Code NVZ
UDI-Device Identifier05414734507585
UDI-Public(01)05414734507585(10)A000039877(17)190531
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Model NumberCD2411-36C
Device Catalogue NumberCD2411-36C
Device Lot NumberA000039877
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2025
Initial Report FDA Received Date11/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/05/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RIATA; TENDRIL SDX
Patient Age72 YR
Patient SexMale
-
-