| Model Number |
A22042A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the resection sheath distal end was broken.The issue occurred during set-up.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updates fields: b5, g3, h2, h3, h6 and h11.The device was returned to olympus for inspection, and the reported failure was confirmed.The most probable cause of the complaint was cause not established, which indicates that the investigation findings do not lead to a clear conclusion about the cause of the reported adverse event.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|