| Model Number |
PC1Q |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during induction, a pc1q while connected to a vitawave plus cuff showed different values compared to an arm cuff.Initially, the finger cuff values were 146/63 (93) while the arm cuff read 165/70 (106).After a few arm adjustments were made and reclipping the pc1q, the finger cuff values were updated to 118/52 (75) and the arm cuff values were 130/55 (80).Patient was undergoing a spine procedure.Both cuffs were not placed on the same arm.It is unknown whether the inaccurate values were caused by the vitawave plus cuff or the pc1q.There were no patient injuries.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional product code: dqe, qaq, mud, dxn, dsb, fll, qms, olw device is not available for return.Without return of the unit, it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.No corrective actions will be taken at this time.Report id: (b)(4) was submitted against the vitawave plus cuff.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
| |
|
Additional Manufacturer Narrative
|
|
This event was found to be a duplicate of an existing report, 2015691-2025-09143.
|
| |
|