| Catalog Number |
EP-115394 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
09/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.D10: concomitant medical products, part (lot): 211216(527760), 211239(357120), 115370(347800), 211228(262520), 211228(262540).Associated product information: (b)(4).The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during a clinical study a patient had an initial right reverse total shoulder arthroplasty.Approximately 15 days post-implantation, the patient complained of pain and received a steroid injection into the deltoid muscle.Posteriorly, approximately 5 years later, persistent pain was reported with no treatment.Finally, approximately 5 years post-implantation, the study was complete with all components remaining implanted.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6; h10.The reported event is confirmed through evaluation of medical records.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: patient experienced post-op aching, pinching pain and minimal stiffness after r-shoulder surgery, managed with pt and injections.Persistent pain noted again in a posterior visit.Mild in severity, treatment: pt prescribed; injections to deltoid administered.Study complete with all components intact if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|