| Model Number |
TSGC0201 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a patient presented with grade 3-4 functional tricuspid regurgitation (tr) for a triclip procedure.During the preparation of steerable guide catheter(sgc), the guide was unable to hold column.A g5 sgc replacement was used to complete the procedure.The final tr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.The investigation was unable to determine a cause for the reported leak (loss of fluid column).There is no indication of a product issue with respect to manufacture, design, or labeling.
|
| |
|