| Model Number |
L300 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded a code 1003, which states that the voltage is too low for projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended as the battery impedance is expected to increase, and radio frequency (rf) telemetry will eventually be disabled.This device remains in-service at this time.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded a code 1003, which states that the voltage is too low for projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended as the battery impedance is expected to increase, and radio frequency (rf) telemetry will eventually be disabled.This device remains in-service at this time.No adverse patient effects were reported.Additional information was received stating that this pacemaker was subsequently explanted and successfully replaced.Other than the intervention no additional adverse patient effects were reported.This device has not been returned for analysis.
|
| |
|