• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO12.6 (-15.00/1.5)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced excessive vaulting, and significant reduction of irido-corneal angles.The lens was exchanged with a shorter length lens and the problem was resolved.The cause of the event was reported as unknown.
 
Additional Manufacturer Narrative
H6 - health effect clinical code: 4581 - significant reduciton of irido-corneal angles.Manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
 
Additional Manufacturer Narrative
Correction: a2 - age at time of event, date of birth: "(b)(6) 2025" corrected to "(b)(6) 2004." b3 - date of event: "29-jun-2004" corrected to "07-jul-2025." claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key23662862
Report Number2023826-2025-01899
Device Sequence Number9082479
Product Code QCB
UDI-Device Identifier00840311322763
UDI-Public(01)00840311322763(11)250618(17)280531
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICMO12.6 (-15.00/1.5)
Device Catalogue NumberVTICMO 12.6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2025
Supplement Date Received by Manufacturer05/01/2026
Initial Report FDA Received Date11/26/2025
Supplement Report FDA Received Date05/04/2026
Date Device Manufactured06/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL-SFC-45, LOT#-UNK.; FOAM TIP PLUNGER MODEL-FTP, LOT#-UNK.; INJECTOR MODEL-MSI-PF, LOT#-UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age21 YR
Patient SexFemale
-
-