| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Head Injury (1879)
|
| Date of Event |
11/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The parent reported a sudden hearing decline with the right ci due to a possible head impact in (b)(6) 2025.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue breakages.Also, an
impact to the head might have weakened the fixation of the active electrode and could also be a factor for electrode mobility.The investigation results appear to match the problems mentioned in the recipient report.This is a final report.
|
| |
|
Event or Problem Description
|
|
The parent reported a sudden hearing decline with the right cochlea implant due to a possible head impact.The user has been re-implanted.
|
| |
|