• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751765
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that poor cutting and aspiration was good during vitrectomy surgery.The surgery was completed after replacement of pack.There was no patient impact.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.2., h.3., h.6.And h.11.No technical inquiries were received from the customer.One opened probe with tip protector in a tray was received for the report.The sample was visually inspected and found to be conforming.The sample was then functionally tested for actuation and cut.The sample was found to be conforming for both tests.The probe engine was disassembled, and the components were inspected.The diaphragm, spacer, top o-ring are all intact.Top o-ring has inner diameter damage.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification (rfid) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation concluded that the root cause of the reported event is not determined as the returned sample was conforming for all visual and functional testing.However, a manufacturing related potential contributor factor was identified, damaged was observed on the o-ring and cannot be ruled out for the actuation problem as reported.The returned sample functionally met specifications: however, a non-conformance was completed for the damage observed on the o-ring.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23664027
Report Number1644019-2025-05102
Device Sequence Number21058087
Product Code LRO
UDI-Device Identifier00380657517657
UDI-Public(01)00380657517657(17)270131(11)250205(10)1782HJ
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065751765
Device Lot Number1782HJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/04/2025
Supplement Date Received by Manufacturer03/17/2026
Initial Report FDA Received Date11/27/2025
Supplement Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/05/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-