| Catalog Number |
72290037 |
| Medical Device Problem Codes |
Device Alarm System (1012); Self-Activation or Keying (1557)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a knee arthroscopy, the button of the werewolf wand was released but it continued to be activated until the system faulted and an error message appeared telling to shut down the console and restart.The procedure was resumed, without any delay, using an equivalent sn back-up.No further complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a knee arthroscopy, the button of the werewolf wand was released but it continued to be activated until the system faulted and an error message appeared telling to shut down the console and restart.It also caused the machine to alarm until the wand was unplugged.The procedure was resumed, without any delay, using an equivalent sn back-up.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h2: additional information in b5 and h6: medical device problem code.
|
| |
|