• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SINGAPORE OPERATIONS LINQ II¿ INSERTABLE CARDIAC MONITOR; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITH ARRHYTHMIA DETECTION)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SINGAPORE OPERATIONS LINQ II¿ INSERTABLE CARDIAC MONITOR; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITH ARRHYTHMIA DETECTION) Back to Search Results
Model Number LNQ22
Medical Device Problem Codes Reset Problem (3019); Data Problem (3196)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an implantable cardiac monitor underwent a hard electrical reset, which resulted in corrupted device counters and a reset of the device clock.The patient had a cardioversion around this time.The patient was in the clinic, and the healthcare provider installed the ramware and corrected the device date and time.The device was expected to function as intended following reprogramming.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
B3: date is approximate.Month and year are confirmed valid.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a full electrical reset occurred.Analysis of the device memory indicated that the device clock was incorrect.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Correction: updated coding in h6.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINQ II¿ INSERTABLE CARDIAC MONITOR
Common Device Name
RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITH ARRHYTHMIA DETECTION)
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN  486056
Manufacturer (Section G)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN   486056
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23666869
Report Number3008973940-2025-09009
Device Sequence Number10431324
Product Code MXD
UDI-Device Identifier00763000060374
UDI-Public(01)00763000060374(17)220408
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K200795
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/08/2022
Device Model NumberLNQ22
Device Catalogue NumberLNQ22
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2025
Supplement Date Received by Manufacturer12/02/2025
02/06/2026
Initial Report FDA Received Date11/27/2025
Supplement Report FDA Received Date12/12/2025
02/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
Patient Weight59 KG
-
-