| Catalog Number |
72205882 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
11/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl procedure, it was noticed that the q-fix knotless anchor's shuttling process of the repair suture, failed.The procedure was resumed after a 15 minute delay, using a similar s+n back up product, using the original drilled bone hole by drilling over it, leaving no void in the patient.Patient is healthy and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: corrected information: d3 and section g: contact office, manufacturing site.This report was inadvertently submitted under manufacturer report number (b)(4), correct manufacturer report number is (b)(4).
|
| |
|
Search Alerts/Recalls
|