• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON TORIC ASPHERIC HYDROPHOBIC ACRYLIC IOL WITH AUTONOME DELIVERY SYSTEM; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON TORIC ASPHERIC HYDROPHOBIC ACRYLIC IOL WITH AUTONOME DELIVERY SYSTEM; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CNA0T3
Medical Device Problem Codes Device Damaged Prior to Use (2284); Failure to Advance (2524)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint "injector seemed non-compliant upon initial opening, plunger override" was not observed, straight leading haptic was observed.The device was received in an opened blister tray inside the product carton.Solution is dried in the device.The lock-out assembly has been removed.The plunger and the lens have been advanced into the mid-nozzle.The plunger position is acceptable.The trailing haptic is folded onto the optic.The leading haptic is extended straight.The reported complaint "injector seemed non-compliant upon initial opening, plunger override" was not observed, straight leading haptic was observed.The plunger, lens and haptic positions are acceptable.The presentation of straight-leading haptics is addressed in the product instructions for use (ifu), where it is provided that delivery may proceed with caution and should be managed based on the outlined instructions.While straight leading haptics may occasionally occur, close adherence to the ifu provides the best opportunity for optimal delivery.Possible associated factors may also include: ¿use of a non-qualified viscoelastic or bss (balanced salt solution) in the delivery system.Material properties of non-qualified ovds (ophthalmic viscosurgical device) may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.¿excessive delay between device preparation and subsequent delivery resulting in the potential for haptic unfolding.¿use of the delivery system at operating room temperatures outside of the recommended range of 18 °c (64 °f) to 23 °c (73 °f).Ovd should be allowed to come to operating room temperature over a 20 minute timeframe prior to use.Any of the above factors alone, or in combination, may led to an event such as reported.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A pharmacist reported that during an intraocular lens (iol) implant procedure, injector seemed non-compliant upon initial opening, plunger override.The surgery was completed the same day after replacing with another injector.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON TORIC ASPHERIC HYDROPHOBIC ACRYLIC IOL WITH AUTONOME DELIVERY SYSTEM
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23671938
Report Number9612169-2025-02383
Device Sequence Number9881534
Product Code MJP
UDI-Device Identifier00380652398794
UDI-Public(01)00380652398794(17)290506(11)250508(10)26020657
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Pharmacist
Type of Report Initial
Report Date (Section B) 11/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNA0T3
Device Lot Number26020657
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/05/2025
Initial Report FDA Received Date11/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
-
-