Internal report reference number: (b)(4).Meter was not returned for evaluation.Test strips were returned for evaluation.Product testing was performed and defect found on returned test strips: high blood readings.Importer's (thi) retention testing was performed using test strips from the same lot.Retention strips tested within specifications.Retention testing was performed by the manufacturer (sinocare inc.) using test strips from the same lot.Sinocare retention strip lot tested within specification.Root cause: (b)(4): storage outside specifications.Notes: 1.Manufacturer contacted customer in several follow-up calls to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.Note 2.In multiple fields the form indicates that manufacturer received product, however, (b)(4) received the product returned and conducted the product evaluation.Note 3.Sinocare (a foreign manufacturer, fei #(b)(4)) and (b)(4).(a u.S.Company/importer, fei #(b)(4)) have entered into a customer service agreement.(b)(4).Handles customer complaints for the trueness¿ blood glucose monitoring system and trueness¿ air blood glucose monitoring system ((b)(4) - class 2) and records customer information, establishing a unique complaint number.
|