• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. NIM® 3.0 INTERFACE; STIMULATOR, NERVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. NIM® 3.0 INTERFACE; STIMULATOR, NERVE Back to Search Results
Model Number 8253200
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that halfway through the surgery, after stimulation with the probe, there was no response (negative or positive).This had happened before.The error persists even after changing the probe.It could be the command fuses.The procedure was completed with reported product.The procedure was completed without monitorization.The user does not confirms there was something causing a lack of response (i.E.Anesthesia, muscle relaxant on board, or too much fluid in the field.There was no patient impact in this event.
 
Additional Manufacturer Narrative
H3: the patient interface is contaminated inside; it can affect the pcb.The cable is discolored.One fuse housing is worn and dented at the outside.One clamp is worn and the other clamp is missing.H6: codes updated.Previous applied fdm b21, fdr c21, fdc d16, img g02030 codes are not applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® 3.0 INTERFACE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key23677451
Report Number1045254-2025-02653
Device Sequence Number12973140
Product Code ETN
UDI-Device Identifier00613994499615
UDI-Public(01)00613994499615(11)100830
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K083124
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8253200
Device Catalogue Number8253200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Supplement Date Received by Manufacturer01/12/2026
Initial Report FDA Received Date11/30/2025
Supplement Report FDA Received Date01/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/30/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-