Additional information provided in d.9., h.3., h.6.And h.11.The two returned product were visually inspected.Visual inspection confirmed that the aspiration line of the probe manifold was bent at the tubing insertion to the probe engine on one of the two samples.The defect prevented fluid from flowing from the probe to the cassette.The investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.The most likely root cause for the customer¿s event is a manufacturing defect.The vitrectomy probe has a rear shell meant to improve the ergonomics of the handpiece, however, if the rear shell is improperly installed, it will result in the customers reported event.Action will not be taken based on this occurrence.An analysis of similar complaints confirmed no unfavorable trend for this event.Routine training is performed to ensure non-conforming product is rejected and placed in a reject bin during production.Quality assurance will continue to monitor and will take action for future occurrences as is deemed necessary.Consumables manufacturing has been made aware of the issue through the monthly complaint review meeting.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|