Model Number 6550202
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having posterior lumbar interbody fusion (l4) due to spondylolisthesis.It was reported that after cement injection, the guide tip remained when the delivery device was removed.They tried to remove it by reattaching the kocher and delivery guide, but extraction was not possible.Due to the pps incision, direct inspection near the screw head revealed cement head leakage.No leakage was observed in the crown area.A lump of cement was found around the position of the screw head and tab (according to the physician¿s initial observation).The cement was broken with a mallet and chisel, and the guide tip was successfully removed.It was confirmed that no cement remained around the screw or crown, and the operation was completed with final locking.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3: product analysis #834894325, part # 6550202: lot # k25d6016, visual and optical inspection confirmed the fns tip is filled with dried cement.It appears the tip may have not been seated fully in the screw head during the attempted cement delivery process which possibly contributed to the cement leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOYAGER FENESTRATED SCREWS
Common Device Name
DISPENSER, CEMENT
Manufacturer (Section D)
WARSAW ORTHOPEDIC, INC.
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDIC, INC.
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key23677903
Report Number1824199-2025-00092
Device Sequence Number8927221
Product Code KIH
UDI-Device Identifier20763000211039
UDI-Public20763000211039
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6550202
Device Catalogue Number6550202
Device Lot NumberK25D6016
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Supplement Date Received by Manufacturer12/17/2025
Initial Report FDA Received Date12/01/2025
Supplement Report FDA Received Date01/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-